Title of article :
A Simple Method of Triggering Balloon Counterpulsation Accurately During Off-Pump Coronary Artery Bypass Surgery
Author/Authors :
Takayuki Ono، نويسنده , , Teiji Asakura، نويسنده , , Takeki Ohashi، نويسنده , , Nagara Ono، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
3
From page :
723
To page :
725
Abstract :
With increasing experience, off-pump coronary artery bypass grafting for high-risk patients can be performed safely. However, in patients who need intraaortic balloon counterpulsation support, mistriggering of intraaortic balloon counterpulsation during mobilization of the heart can induce unstable hemodynamic conditions. My colleagues and I have developed a simple method of detecting the trigger signal accurately: an epicardial pacemaker wire is placed close to the apex of the left ventricle, and 1 precordial V lead is disconnected and then linked to the epicardial pacemaker wire. This method provides an excellent detection of R-wave potentials in any position of the heart throughout an entire off-pump coronary artery bypass grafting procedure.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2005
Journal title :
The Annals of Thoracic Surgery
Record number :
608364
Link To Document :
بازگشت