• Title of article

    Automatic Intraaortic Balloon Pump Timing Using an Intrabeat Dicrotic Notch Prediction Algorithm

  • Author/Authors

    Jan J. Schreuder، نويسنده , , Alessandro Castiglioni، نويسنده , , Andrea Donelli، نويسنده , , Francesco Maisano، نويسنده , , Jos R.C. Jansen، نويسنده , , Ramzi Hanania، نويسنده , , Pat Hanlon، نويسنده , , Jan Bovelander، نويسنده , , Ottavio Alfieri، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    1017
  • To page
    1022
  • Abstract
    Purpose The efficacy of intraaortic balloon counterpulsation (IABP) during arrhythmic episodes is questionable. A novel algorithm for intrabeat prediction of the dicrotic notch was used for real time IABP inflation timing control. Description A windkessel model algorithm was used to calculate real-time aortic flow from aortic pressure. The dicrotic notch was predicted using a percentage of calculated peak flow. Automatic inflation timing was set at intrabeat predicted dicrotic notch and was combined with automatic IAB deflation. Evaluation Prophylactic IABP was applied in 27 patients with low ejection fraction (< 35%) undergoing cardiac surgery. Analysis of IABP at a 1:4 ratio revealed that IAB inflation occurred at a mean of 0.6 ± 5 ms from the dicrotic notch. In all patients accurate automatic timing at a 1:1 assist ratio was performed. Seventeen patients had episodes of severe arrhythmia, the novel IABP inflation algorithm accurately assisted 318 of 320 arrhythmic beats at a 1:1 ratio. Conclusions The novel real-time intrabeat IABP inflation timing algorithm performed accurately in all patients during both regular rhythms and severe arrhythmia, allowing fully automatic intrabeat IABP timing.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    2005
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    608428