Title of article :
Complete Atrioventricular Block Decreases Left Ventricular Assist Device Flow Rate
Author/Authors :
Frederic Collart، نويسنده , , Pierre Dieuzaide، نويسنده , , Francois Kerbaul، نويسنده , , Annick Mouly-Bandini، نويسنده , , Thierry G. Mesana، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
A 52-year-old man, living at home, had been on left ventricular assistance for 15 months with a Novacor pump (Worldheart, Ottawa, ON, Canada). Following the onset of effort dyspnea that had recently become worse, he was admitted to the hospital. A review of the pump showed that its flow rate had significantly decreased. An electrocardiogram showed a third-degree atrioventricular block. A pacemaker was implanted and the patient became asymptomatic immediately afterwards. This clinical case report shows the importance of satisfactory right ventricle filling in these patients. Right ventricle filling can be impaired by conduction disorders.
Journal title :
The Annals of Thoracic Surgery
Journal title :
The Annals of Thoracic Surgery