Title of article :
Intraoperative Effects of the Coapsys Annuloplasty System in a Randomized Evaluation (RESTOR-MV) of Functional Ischemic Mitral Regurgitation
Author/Authors :
Eugene A. Grossi، نويسنده , , Paul C. Saunders، نويسنده , , Y. Joseph Woo، نويسنده , , Deepak M. Gangahar، نويسنده , , John C. Laschinger، نويسنده , , David C. Kress، نويسنده , , Michael P. Caskey، نويسنده , , Charles F. Schwartz، نويسنده , , James Wudel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
6
From page :
1706
To page :
1711
Abstract :
Background Functional ischemic mitral regurgitation (MR) frequently arises after myocardial infarction; it is characterized by annular enlargement or lateral displacement of the subvalvular apparatus. Coapsys is a ventricular–annular remodeling device designed to treat functional ischemic MR; it does not require cardiopulmonary bypass. Initial intraoperative results of the RESTOR-MV randomized clinical trial are presented. Methods Patients referred for coronary artery bypass grafting with preoperative MR grade of 2 or greater were studied, excluding those with structural valve abnormalities. The Coapsys device, which consists of two epicardial pads connected by a flexible cord, was surgically implanted in 19 patients. Under epicardial echocardiographic guidance, the cord was passed through the left ventricle and tightened externally to improve leaflet coaptation and stabilize the ventricular wall; tightening was conducted with color flow Doppler imaging. Results Patients were 64.5 ± 9.2 years old with an ejection fraction of 0.383 ± 0.089 and received 2.7 ± 1.1 grafts. Intraoperative MR grade was 2.7 ± 0.8 after induction and was reduced to 0.4 ± 0.7 after implantation (p< 0.0001). Mean epicardial dimension was reduced from 8.5 ± 1.2 to 6.4 ± 0.9 cm (p< 0.0001). Intraoperative MR was reduced in 95% (18 of 19) of patients, and 84% (16 of 19) had MR grade 1 or less after implantation. All implants were performed without cardiopulmonary bypass or conversion to standard annuloplasty. No hemodynamic compromise or structural damage to the mitral apparatus was noted. Significant acute remodeling was noted in the left ventricular dimensions. Conclusions In patients without structural valve disease, the Coapsys device acutely reduces functional MR. Further randomized evaluation will assess long-term stability and compare it with standard annuloplasty techniques.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2005
Journal title :
The Annals of Thoracic Surgery
Record number :
609118
Link To Document :
بازگشت