Title of article
Six-Month Angiographic Follow-Up of the PAS-Port II Clinical Trial
Author/Authors
Jan F. Gummert، نويسنده , , Stefanos Demertzis، نويسنده , , Klaus Matschke، نويسنده , , Utz Kappert، نويسنده , , Marcel Anssar، نويسنده , , Francesco Siclari، نويسنده , , Volkmar Falk، نويسنده , , Edwin L. Alderman، نويسنده , , Wolfgang Harringer، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
7
From page
90
To page
96
Abstract
Background
The PAS-Port device (Cardica, Redwood City, CA) allows the rapid deployment of a clampless proximal anastomosis between a vein graft and the aorta.
Methods
Fifty-four patients awaiting elective coronary artery bypass graft surgery were enrolled. Outcome variables were intraoperative device performance, early and 6- month angiographic graft patency, and 12-month clinical follow-up.
Results
Sixty-three PAS-Port devices were deployed in 54 patients. Two deployments were unsuccessful. There were no reoperations for bleeding. Two patients died of causes unrelated to the device. Patency evaluation at discharge was performed by angiogram on 49 implants and computed tomography in 2 implants (86% follow-up). At discharge, all evaluated grafts were patent (100%) and rated Fitzgibbon A. At 6-month follow-up, there was no additional mortality; 47 implants (88% follow-up) were evaluated by angiography (Fitzgibbon O [n = 1], Fitzgibbon B [n = 1], and Fitzgibbon A [n = 45]) and 5 by computed tomography. All grafts but 1 were patent (98.1%). At 12 months, 2 additional patients died of causes unrelated to the PAS-Port implant. Forty-six of 50 alive patients (95.8%) were followed up without any reports of device-related major adverse cardiac events.
Conclusions
Discharge (100%) and 6-month patency (98%) are excellent; patency and 12 months’ clinical follow-up compares favorably with data from historical hand-sewn controls. The PAS-Port system safely allows the clampless creation of a proximal anastomosis.
Background
The PAS-Port device (Cardica, Redwood City, CA) allows the rapid deployment of a clampless proximal anastomosis between a vein graft and the aorta.
Methods
Fifty-four patients awaiting elective coronary artery bypass graft surgery were enrolled. Outcome variables were intraoperative device performance, early and 6- month angiographic graft patency, and 12-month clinical follow-up.
Results
Sixty-three PAS-Port devices were deployed in 54 patients. Two deployments were unsuccessful. There were no reoperations for bleeding. Two patients died of causes unrelated to the device. Patency evaluation at discharge was performed by angiogram on 49 implants and computed tomography in 2 implants (86% follow-up). At discharge, all evaluated grafts were patent (100%) and rated Fitzgibbon A. At 6-month follow-up, there was no additional mortality; 47 implants (88% follow-up) were evaluated by angiography (Fitzgibbon O [n = 1], Fitzgibbon B [n = 1], and Fitzgibbon A [n = 45]) and 5 by computed tomography. All grafts but 1 were patent (98.1%). At 12 months, 2 additional patients died of causes unrelated to the PAS-Port implant. Forty-six of 50 alive patients (95.8%) were followed up without any reports of device-related major adverse cardiac events.
Conclusions
Discharge (100%) and 6-month patency (98%) are excellent; patency and 12 months’ clinical follow-up compares favorably with data from historical hand-sewn controls. The PAS-Port system safely allows the clampless creation of a proximal anastomosis.
Journal title
The Annals of Thoracic Surgery
Serial Year
2006
Journal title
The Annals of Thoracic Surgery
Record number
609285