Title of article :
Fracture of Colvin-Galloway Future Band Causing a Tear in the Anterior Mitral Leaflet
Author/Authors :
Martin Hartrumpf، نويسنده , , Ulrich A. Stock، نويسنده , , Ralf-Uwe Kuehnel، نويسنده , , Johannes M. Albes، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
2
From page :
1879
To page :
1880
Abstract :
The first case of a fracture of a semi-rigid annuloplasty device is reported. A 56-year-old man presented with recurrence of severe mitral and tricuspid regurgitation 10 months after proper implantation of a Colvin-Galloway Future band (Medtronic Inc, Minneapolis, MN). During redo surgery we made the unexpected finding that its metal skeleton was fractured near the posterior segment. A tear in the anterior mitral leaflet was also found. It is speculated that the device fracture may have put additional strain on the native annulus.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2006
Journal title :
The Annals of Thoracic Surgery
Record number :
609660
Link To Document :
بازگشت