Title of article :
Leakage of the Arterial Prosthesis of an Impella RVAD
Author/Authors :
Hendrik T. Tevaearai، نويسنده , , Juerg Schmidli، نويسنده , , Paul Mohacsi، نويسنده , , Hans-Ulrich Rothen، نويسنده , , Friedrich S. Eckstein، نويسنده , , Thierry P. Carrel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
3
From page :
1527
To page :
1529
Abstract :
Seromas occurring around a vascular graft are a rare complication. We report a life-threatening plasma leakage that occurred through the polytetrafluoroethylene vascular prosthesis of an Impella right ventricular assist device (Impella RD [Impella Cardiosystems GmbH, Aachen, Germany]) implanted in a 62-year-old patient with acute right ventricular failure after cardiac transplantation. The leakage became progressively massive. Weaning the patient from the right ventricular assist device was not possible. The prosthesis was thus wrapped within a pericardial patch to contain the leakage. Three days later the patient could be successfully weaned and the pump was removed. The clinical evolution was favorable.