• Title of article

    Composite stentless xenograft for aortic valve replacement: Clinical evaluation of function

  • Author/Authors

    Mark F. OʹBrien، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    4
  • From page
    406
  • To page
    409
  • Abstract
    At The Prince Charles Hospital, 55 patients have received the composite porcine aortic xenograft (model 300, Cryolife-OʹBrien valve; Cryolite International, Atlanta, GA; formerly the Bravo Cardiovascular Inc valve) for aortic valve replacement since December 1992. Associated procedures were required in 50% of patients (coronary artery bypass grafting in 25 patients). Hospital and “late” mortality have been 0%. Follow-up has been 100% complete. The median patient age was 74 years (range, 59 to 88 years). One permanent intraoperative cerebrovascular accident and two early thromboembolic events (both associated with acute atrial fibrillation) have constituted the only three morbid events. Serial echocardiography at 1 week, 6 months, and 1 year has shown a mean gradient of 9 mm Hg. A small number of patients have a “trivial” degree of valve incompetence. No progression of either gradient or incompetence is evident. The clinical state of 54 of these patients is most satisfactory. However, longterm surveillance is necessary to determine the 8- to 10-year durability of this stentless xenograft.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    1995
  • Journal title
    The Annals of Thoracic Surgery
  • Record number