Title of article :
Modified method for novacor left ventricular assist device implantation
Author/Authors :
Mario Viganoʹ، نويسنده , , Luigi Martinelli، نويسنده , , Gaetano Minzioni، نويسنده , , Mauro Rinaldi، نويسنده , , Francesco Pagani، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
3
From page :
247
To page :
249
Abstract :
We have developed and applied in 7 cases an alternative “orthodromic” technique for Novacor left ventricular assist device implantation. After sternotomy, the pocket is tailored, dividing the insertion of the diaphragm to the lower ribs, for easier bleeding control. On cardiopulmonary bypass, the aorta is cress-clamped and cardioplegia administered. Apical cannulation is performed first on a dry, still field. The device is then easily deaired, with blood flowing in the physiologic direction. The aorta is declamped and the outflow conduit is anastomosed. Before the suture is tied, the final deairing is obtained. This technique allows extreme precision in apical cannulation, easier control of bleeding, and accurate deairing of the pump. The ischemic time is short and damage to the right ventricle negligible.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
1996
Journal title :
The Annals of Thoracic Surgery
Record number :
613148