Title of article
Cardiac pacemaker dysfunction in children after thoracic drainage catheter manipulation
Author/Authors
Kevin W. Lobdell، نويسنده , , Henry L. Waiters III، نويسنده , , Claudia Hudson، نويسنده , , Mehdi Hakimi، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
2
From page
1466
To page
1467
Abstract
Two children underwent placement of permanent, epicardial-lead, dual-chamber, unipolar pacemaker systems for complete heart block. Postoperatively, both patients demonstrated subcutaneous emphysema—in the area of their pulse generators—temporally related to thoracic catheter manipulation. Acutely, each situation was managed with manual compression of the pulse generator, ascertaining appropriate pacemaker sensing and pacing. Maintenance of compression with pressure dressings, vigilant observation/monitoring, and education of the care givers resulted in satisfactory pacemaker function without invasive intervention.
Journal title
The Annals of Thoracic Surgery
Serial Year
1997
Journal title
The Annals of Thoracic Surgery
Record number
614242
Link To Document