Title of article :
Discrepancies Between Sizer and Valve Dimensions: Implications for Small Aortic Root
Author/Authors :
Claus Bartels MD، نويسنده , , Rainer G. Leyh، نويسنده , , J. F. Matthias Bechtel، نويسنده , , Elrina Joubert-Hübner، نويسنده , , Hans-Hinrich Sievers MD، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
Background. Precise labeling of sizer and valve diameters is crucial for optimal valve selection especially in the small aortic root. This study examines the accuracy of manufacturers’ markings on small aortic prostheses and sizers.
Methods. Sizer and valve dimensions of 22 different mechanical aortic prostheses (19 to 23 mm) were evaluated by caliper micrometer measurements.
Results. Nearly all sizers exceeded their marked dimensions by up to 1.0 mm. Measured tissue annulus diameters for 19-mm-labeled valves varied between 18.3 and 19.6 mm, for 21-mm valves from 20.5 to 21.6 mm, and for 23-mm valves from 22.4 to 23.5 mm, respectively. The orifice areas ranged from 1.5 to 2.06 cm2 for 19-mm valves, from 2.0 to 2.55 cm2 for 21-mm valves, and from 2.4 to 3.09 cm2 for 23-mm valves, respectively.
Conclusions. Actual sizer dimensions and tissue annulus diameters of various small mechanical aortic prostheses varied considerably from their marked diameters. These differences should be considered to ensure the optimal prosthesis selection for each patient.
Journal title :
The Annals of Thoracic Surgery
Journal title :
The Annals of Thoracic Surgery