Title of article
Hemodynamic features of the freestyle aortic bioprosthesis compared with stented bioprosthesis
Author/Authors
Jean G. Dumesnil، نويسنده , , Marie-Hélène LeBlanc، نويسنده , , Paul C. Cartier، نويسنده , , Jacques Métras، نويسنده , , Denis Desaulniers، نويسنده , , Daniel P. Doyle، نويسنده , , Michel D. Lemieux، نويسنده , , Gilles Raymond، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
4
From page
130
To page
133
Abstract
Background. The Freestyle prosthesis is a new stentless aortic bioprosthesis. Anticipated benefits are improved hemodynamics and increased longevity.
Methods. Doppler echocardiograms were performed early and at 3 to 6 months, 1 year, and 2 years after operation in 157 patients (69 men, 88 women, aged 48 to 85 years) with this prosthesis, and results were compared with hemodynamic data in patients with Intact and Mosaic stented bioprostheses.
Results. Distinctive features of the prosthesis compared with stented prostheses are (1) an increase in effective orifice area (+0.15 ± 0.26 cm2; p< 0.05) and a decrease in mean gradient (−3.5 ± 4.0 mm Hg; p< 0.001) during the first 3 to 6 months postoperatively and stabilization thereafter; (2) a markedly lower mean gradient at 1 year after operation (average, 6 ± 4 mm Hg) than in stented prostheses (Intact, 22 ± 8 mm Hg; Mosaic, 12 ± 6 mm Hg); (3) in contrast to stented prostheses, in vivo effective orifice areas much lower (−0.91 ± 0.35 cm2) than those calculated in vitro; (4) as in stented prostheses, the indexed effective orifice area (cm2/m2) is the best predictor (r = 0.77 at 1 year) of the mean gradient after operation; and (5) similar incidence of aortic regurgitation (trivial or mild, 34% versus 29% in Intact).
Conclusions. The hemodynamics of the Freestyle are very satisfactory and represent a marked improvement in comparison to stented prosthesis
Journal title
The Annals of Thoracic Surgery
Serial Year
1998
Journal title
The Annals of Thoracic Surgery
Record number
615525
Link To Document