Title of article
Clinical and hemodynamic performance of the freestyle aortic root bioprosthesis
Author/Authors
Paul C. Cartier، نويسنده , , Jean G. Dumesnil، نويسنده , , Jacques Métras، نويسنده , , DenisDesaulniers، نويسنده , , Daniel P. Doyle، نويسنده , , Michel D. Lemieux، نويسنده , , Gilles Raymond، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
5
From page
345
To page
349
Abstract
Background. The objective of this study is to assess the clinical and hemodynamic performance of a stentless porcine bioprosthesis, the Freestyle aortic root bioprosthesis.
Methods. Consenting patients requiring isolated aortic valve or aortic root replacement received the Freestyle bioprosthesis. Clinical follow-up and echocardiographic data were obtained at discharge, 3 to 6 months, 1 year, and annually thereafter.
Results. Two hundred seventy-six patients received a Freestyle aortic root bioprosthesis between January 1993 and July 1997. The mean age was 67.7 years. Preoperatively, 86.3% were either New York Heart Association class III or IV. Two hundred thirty-eight patients underwent valve (subcoronary) replacement, 36 underwent aortic root replacement, and 2 underwent valve replacement using the root-inclusion technique. The early mortality was 5.4%, with 3.3% mortality for the subcoronary technique and 19.4% mortality for aortic root replacement. The mean gradient decreased significantly between discharge and the 3- to 6-month follow-up and stabilized thereafter. The effective orifice area increased significantly from discharge to 3 to 6 months’ follow-up. At 3 years, 84.4% of the patients had either no or trivial regurgitation.
Conclusions. The Freestyle bioprosthesis has good clinical performance and good short-term hemodynamic performance. The majority of the regurgitation identified is not clinically significant.
Journal title
The Annals of Thoracic Surgery
Serial Year
1999
Journal title
The Annals of Thoracic Surgery
Record number
615682
Link To Document