Title of article :
Pseudoaneurysm of the outflow graft in a patient with novacor N100 LVAD system
Author/Authors :
Christoph Knosalla، نويسنده , , Yu-guo Weng، نويسنده , , Semih Buz، نويسنده , , Matthias Loebe، نويسنده , , Roland Hetzer، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
3
From page :
1594
To page :
1596
Abstract :
Six months after implantation of a wearable Novacor N100 left ventricular assist device, a 47-year-old patient developed a swelling that overlay the body of sternum. Computed tomographic scan of the chest revealed a pseudoaneurysm of the Novacor outflow graft. The patient was taken back to surgery and the diagnosis was confirmed at operation. Repair was performed by direct sutures via right anterolateral thoracotomy, under deep hypothermia and low flow technique. The postoperative course was uneventful.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2000
Journal title :
The Annals of Thoracic Surgery
Record number :
616808
Link To Document :
بازگشت