Title of article :
A device for real-time, intraoperative margin assessment in breast-conservation surgery
Author/Authors :
Tami Karni، نويسنده , , Itzhak Pappo، نويسنده , , Judith Sandbank، نويسنده , , Oleg Lavon، نويسنده , , Varda Kent، نويسنده , , Rona Spector، نويسنده , , Sara Morgenstern، نويسنده , , Shlomo Lelcuk، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
7
From page :
467
To page :
473
Abstract :
Background This trial was designed to study performance of a novel handheld probe (Dune Medical Devices, Caesarea, Israel) in intraoperative detection of positive margins and its potential benefit toward minimizing the positive margin rate. Methods The probe was intraoperatively applied to 57 lumpectomy specimens. Surgeons were blinded to device output, and surgical decisions were not affected by probe data. Probe readings were compared with histological analysis per margin and per patient. Results Nineteen of 22 (86%) pathology-positive patients were intraoperatively detected with device use. Per-margin sensitivity was .71, and specificity was .68, maintained within a range of positive margin definitions (0–.4 cm). Conclusions The device is an effective tool for intraoperative detection of positive margins with the potential for significant positive margin rate reduction.
Keywords :
Intraoperative margin assessment , surgical margins , Lumpectomy , breast cancer , Breast-conservation surgery
Journal title :
The American Journal of Surgery
Serial Year :
2007
Journal title :
The American Journal of Surgery
Record number :
618837