• Title of article

    Effect of Axial Length on Laer pot ize During Photodynamic Therapy: An Experimental tudy in Monkey

  • Author/Authors

    Mineo Kondo، نويسنده , , Yauki Ito، نويسنده , , Kentaro Miyata، نويسنده , , Nagako Kondo، نويسنده , , Kohei Ihikawa، نويسنده , , Hiroko Teraaki، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    2
  • From page
    214
  • To page
    215
  • Abstract
    Purpoe To invetigate the effect of horter axial length on the laer pot ize and laer energy during photodynamic therapy (PDT) in monkey. Deign Experimental tudy with four rheu monkey. Method PDT wa performed on the normal retina of monkey whoe ocular axial length are horter (19.55 to 20.25 mm) than that of human. After the PDT, the eye were enucleated, and the diameter of the irradiated laer pot wa meaured with a microcaliper. Reult The area of actual laer pot wa only 0.56 to 0.61 time of the planned area, which indicated that the laer energy/area wa 1.64 to 1.78 time more intene than planned initially. Concluion Thee reult are the in vivo demontration that the diameter of PDT laer pot i maller for eye with horter axial length.
  • Journal title
    American Journal of Ophthalmology
  • Serial Year
    2006
  • Journal title
    American Journal of Ophthalmology
  • Record number

    626039