Title of article :
Agreement of Rebound Tonometer in Meauring Intraocular Preure With Three Type of Applanation Tonometer
Author/Authors :
Makoto Nakamura، نويسنده , , Urtogtah Darhad، نويسنده , , Yauko Tatumi، نويسنده , , Miyuki Fujioka، نويسنده , , Azua Kuuhara، نويسنده , , Hidetaka Maeda، نويسنده , , Akira Negi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
3
From page :
332
To page :
334
Abstract :
Purpoe To evaluate the agreement of iCare rebound tonometer in meauring intraocular preure (IOP) with Goldmann applanation tonometer (GAT), Tonopen XL, and noncontact tonometer, and the influence of the central corneal thickne (CCT) on IOP meaurement made with thee four tonometer in 45 (12 control and 33 glaucomatou or ocular hypertenive) eye. Deign Clinically relevant experimental tudy. Method Tonometer intermethod agreement wa aeed by the Bland-Altman method. The relation of CCT with abolute IOP value and intertonometer difference were analyzed by linear regreion. Reult The mean difference (95% limit of agreement) in IOP reading between iCare and GAT, Tonopen XL, and noncontact tonometer were 1.40 ± 4.29, 0.00 ± 4.78, and 2.22 ± 4.19 mm Hg, repectively. All tonometrie had a marked aociation with CCT. A the CCT got thicker, iCare coniderably overetimated GAT and Tonopen XL. Concluion Although influenced by CCT, iCare agree well with applanation tonometer.
Journal title :
American Journal of Ophthalmology
Serial Year :
2006
Journal title :
American Journal of Ophthalmology
Record number :
626428