Title of article :
Resistance to the migration of microorganisms of a needle-free disinfectable connector
Author/Authors :
Juan C. Yébenes، نويسنده , , Rafael Mart?nez، نويسنده , , Mateu Serra-Prat، نويسنده , , Goretti Sauca، نويسنده , , Josep A. Capdevila، نويسنده , , Xavier Balanz?، نويسنده , , Mercedes Palomar، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
3
From page :
462
To page :
464
Abstract :
The aim of this study was to assess the efficacy of a disinfectable needle-free connector in reducing the pass of micro-organisms to the lumen of the catheter. A prospective, controlled, experimental trial was performed in which a laboratory model simulated the insertion of a peripheral venous catheter. Catheters inserted in sterile conditions in a hemoculture bottle were closed with the disinfectable needle-free connector (study group) or with a cap (control group). After 9 days of contamination and manipulation of the connector and cap external surfaces, 100% of bottles in the control group were contaminated whereas 60% remained sterile in the study group. The disinfectable needle-free connector showed more resistance to the pass of microorganisms than the conventional cap according to our experimental model.
Journal title :
American Journal of Infection Control (AJIC)
Serial Year :
2002
Journal title :
American Journal of Infection Control (AJIC)
Record number :
635641
Link To Document :
بازگشت