Title of article :
Time-quantified fetal movement detection with two-transducer data fusion,
Author/Authors :
Curtis L. Lowery، نويسنده , , William A. Russell Jr.، نويسنده , , James D. Wilson، نويسنده , , Robert C. Walls، نويسنده , , Pamela Murphy، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
Objective: Our purpose was to develop an automated ultrasound-based fetal movement detection system to batter define fetal movements.
Study design: One hundred one patients had fetal movements recorded over a 20-minute period. Results of movement detection by a single-transducer system (Russell 1) and a two-transducer fusion system (Russell 2) were compared with those of Hewitt-Packard (HP-M-1350-A), maternal perception, and expert ultrasonography review.
Results: A total of 86,592 seconds of videotape was scored for fetal movement, Russell 2 had a second-per-second sensitivity of 57.21% compared with Russell 1 at 40.95%, the HP-M-1350-A at 31.44%, and maternal perception at 30.80%. Russell 2 detected 67.57% of discrete movements compared with 57.52%, 41.98%, and 37.92%, respectively, in other systems.
Conclusion: Russell 2 represents a significant improvement over existing systems in detection of fetal movement on a second-per-second basis.
Keywords :
Fetal body movement , two-transducer system , second-per-second timing , computer-controlled video cassette recorder , Discrete movement
Journal title :
American Journal of Obstetrics and Gynecology
Journal title :
American Journal of Obstetrics and Gynecology