• Title of article

    Complications caused by extramembranous placement of intrauterine pressure catheters,

  • Author/Authors

    Brent K. Lind، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    2
  • From page
    1034
  • To page
    1035
  • Abstract
    A case report is described in which the inadvertent placement of a standard intrauterine pressure catheter in a laboring woman caused partial abruptio placentae and disseminated intravascular coagulation. Altering catheter placement technique and giving attention to aspects of placement can help avoid mishaps, and awareness of possible complications can lead to earlier diagnosis with increased appropriate intervention. (Am J Obstet Gynecol 1999;180:1034-5.)
  • Keywords
    Extramembranous , extraovular placement , intrauterinepressure , intrauterine pressure catheter (IUPC)
  • Journal title
    American Journal of Obstetrics and Gynecology
  • Serial Year
    1999
  • Journal title
    American Journal of Obstetrics and Gynecology
  • Record number