Title of article
Determination of Cr(III) and Cr(VI) in Environmental Waters by Derivative Flame Atomic Absorption Spectrometry Using Flow Injection On-Line Preconcentration with Double-Microcolumn Adsorption
Author/Authors
H-W. Suna، نويسنده , , W-J. Kangb، نويسنده , , J. Hac، نويسنده , , S-X. Lianga and S-G. Shena، نويسنده ,
Issue Information
فصلنامه با شماره پیاپی سال 2004
Pages
7
From page
40
To page
46
Abstract
The species of Cr(III) and Cr(VI) in water samples were determined by flow injection on-line preconcentration and separation on two-microcolumn system–derivative flame atomic absorption spectrometry during a collaborative analysis for certification. The Cr(III) and Cr(VI) in water samples were retained on two microcolumns with ion exchange resin and were eluted directly to nebulizer by 15% HNO3 and 8% NH4NO3, respectively. The characteristic concentration (at the sensitivity grade of 2 mV min-1 for 1 min of preconcentration time) for Cr(III) and Cr(VI) were 0.130 and 0.0985 μg l-1, in the order which were 332- and 431-fold better than those of FAAS, and 45- and 47-fold better than those of FI-FAAS, respectively. The relative standard deviations were 3.27% and 3.66% with corresponding detection limits (3σ) of 0.244 and 0.235 μg l-1, respectively. The linear ranges of determinations for Cr(III) and Cr(VI) were 0~100 μgm l-1 with correlation coefficients of 0.9984 to 0.9996. The satisfactory recovery of 94.4%~106% for Cr(III) and Cr(VI) could be obtained from water samples
Keywords
chromium(III) , Chromium(VI) , Flow injection,On-line preconcentration,Derivative flame atomic absorption spectrometry
Journal title
Journal of the Iranian Chemical Society (JICS)
Serial Year
2004
Journal title
Journal of the Iranian Chemical Society (JICS)
Record number
666429
Link To Document