Title of article :
Determination of Salicylate by Selective and Sensitive Polymer Membrane Electrode: with Internal Solution and Solid Contact
Author/Authors :
Mazloum-Ardakani ، Mohammad نويسنده Department of Chemistry, Faculty of Science, Yazd University, Yazd, I.R. Iran , , Sheikh-Mohseni، Mohammad Ali نويسنده Department of Chemistry, Faculty of Science, Yazd University, Yazd, I.R. Iran , , Benvidi، Ali نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی سال 2010
Abstract :
In this work we developed a new salicylate-selective membrane based on the
trioctylmethyl ammonium chloride as the ion exchanger. The treatment of membrane
investigated and compared in two different types. In one type membrane put on the PVC tube,
electrode with internal solution (IS), and in another type membrane glued on a glassy carbon
electrode, solid contact electrode (SC). Both sensors show a Nernstian response (56±1.0 and
57.7±1.0 mV decade-1, respectively) to salicylate ions over a very wide concentration ranges
(3×10-6?1 M and 1×10-6?1 M, respectively). The detection limits of these sensors are 1×10-6
and 8×10-7 M, respectively. The electrodes possess satisfactory reproducibility, very short
response time (~ 5 s) in the whole concentration ranges, and discriminating ability for
salicylate ions with respect to the most common inorganic anions and special organic and
biological anions. The selectivity coefficients, linear range and detection limit were better for
membrane that glued on glassy carbon electrode. Two electrodes could be used in pH rang 5.0
to 12.0. The electrodes were successfully used for determination of salicylate in real samples.