• Title of article

    Performance of a commercial radon-in-water measurement kit

  • Author/Authors

    Michael E. Kitto، نويسنده , , Eileen M. Fielman، نويسنده , , Douglas K. Haines، نويسنده , , Traci A. Menia، نويسنده , , Abdul Bari، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    3
  • From page
    1255
  • To page
    1257
  • Abstract
    Methods currently approved for the measurement of radon (222Rn) in water in New York State are liquid scintillation counting and emanation into alpha-scintillation cells. A passive system using an electret ion chamber (EIC) was evaluated as an alternative for the measurement of radon in water. Over 130 water samples from a community water supply containing 32 Bq L−1 and 30 standards containing 686 Bq L−1 were measured using the EIC method over 1- to 4-day exposure times. For comparison, identical samples were measured using liquid scintillation counting. Results of duplicate samples were typically within 5% for liquid scintillation counting and within 10% for the EIC. With respect to accuracy, the EIC produced results that were consistently low by 11–15%.
  • Keywords
    RadonGroundwaterElectretLiquid scintillationBias
  • Journal title
    Journal of Environmental Radioactivity
  • Serial Year
    2008
  • Journal title
    Journal of Environmental Radioactivity
  • Record number