Title of article :
An interlaboratory round robin study on the analysis of toxaphene in a cod liver oil standard reference material
Author/Authors :
P. Andrews، نويسنده , , K. Headrick، نويسنده , , J-C. Pilon، نويسنده , , B. Lau، نويسنده , , D. Weber، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
10
From page :
4393
To page :
4402
Abstract :
An inter-laboratory round robin analysis of toxaphene in a National Institute of Standards Cod Liver Oil #1588 was carried out. Analysis was performed by GC-ECD, electron capture negative ion GCMS (ECNI GCMS) and electron impact GC/MS in high resolution single ion monitoring mode (EI-HRSIM). Results of the analyses by GC-ECD, ECNI and HRSIM were 4.22±2.45 ppm, 3.95±1.57 ppm, and 2.35±0.06 ppm respectively. Due to the wide inter-laboratory variation in workup, GC conditions, detection method, and quantitation algorithms used, no one set of factors stood out as the cause of the variation in results.
Journal title :
Chemosphere
Serial Year :
1995
Journal title :
Chemosphere
Record number :
722541
Link To Document :