Title of article
PCB/Toxaphene group separation on silica prior to congener specific determination of toxaphene residues in fish and other samples by GC/ECD
Author/Authors
Bernd Krock، نويسنده , , Walter Vetter، نويسنده , , Bernd Luckas، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
12
From page
1519
To page
1530
Abstract
A precise quantification of toxaphene residues of environmental samples by gas chromatography/ electron capture detection (GC/ECD) requires the separation of the bulk of the polychlorinated biphenyls (PCBs) from the compounds of technical toxaphene (CTT) fraction. For this reason, a PCB/CTT group separation on silica was developed 88-1413 (Parlay #26) and B7-515 (Parlay #32) eluted as first and last out of ten important CTT standards, and can be used to determine the elution volume of the CTT fraction on silica. GC/ECD quantification of CTTs was possible after separation of PCBs on 8.0 g activated silica eluted with 48 mL n-hexane followed by quantitative elution of CTTs with n-hexane/toluene (65:35; v/v). This method is a compromise between separation efficiency and consumption of material. Finally, eight CTTs were quantified in cod liver samples from Iceland and the Baltic Sea.
Journal title
Chemosphere
Serial Year
1997
Journal title
Chemosphere
Record number
723299
Link To Document