• Title of article

    Use of the metastable atom bombardment (MAB) ion source for the elimination of PCDE interference in PCDD/PCDF analysis

  • Author/Authors

    S. Moore، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    5
  • From page
    121
  • To page
    125
  • Abstract
    The analysis of polychlorinated dioxins and furans (PCDDs/PCDFs) can produce erroneous results when polychloro diphenyl ethers (PCDEs) are present because they fragment and rearrange under electron ionization to give isobaric ions with PCDFs. Mass information must be generated to indicate the occurrence of possible PCDEs but uncertainty still exists if a PCDE congener has the same retention time as a PCDF congener. The metastable atom bombardment (MAB) ion source does not transfer enough energy to PCDE upon ionization to fragment the molecular ion, thus eliminating the need for PCDE tracking in PCDF analysis. Experiments were conducted with different gases, representing different ionization energies, to demonstrate that PCDE interference can be eliminated from PCDF analysis.
  • Keywords
    environment , Dioxin and furan analysis , Mass spectrometry , Polychlorinated diphenyl ether , PCDE interference
  • Journal title
    Chemosphere
  • Serial Year
    2002
  • Journal title
    Chemosphere
  • Record number