Title of article
Use of the metastable atom bombardment (MAB) ion source for the elimination of PCDE interference in PCDD/PCDF analysis
Author/Authors
S. Moore، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
5
From page
121
To page
125
Abstract
The analysis of polychlorinated dioxins and furans (PCDDs/PCDFs) can produce erroneous results when polychloro diphenyl ethers (PCDEs) are present because they fragment and rearrange under electron ionization to give isobaric ions with PCDFs. Mass information must be generated to indicate the occurrence of possible PCDEs but uncertainty still exists if a PCDE congener has the same retention time as a PCDF congener. The metastable atom bombardment (MAB) ion source does not transfer enough energy to PCDE upon ionization to fragment the molecular ion, thus eliminating the need for PCDE tracking in PCDF analysis. Experiments were conducted with different gases, representing different ionization energies, to demonstrate that PCDE interference can be eliminated from PCDF analysis.
Keywords
environment , Dioxin and furan analysis , Mass spectrometry , Polychlorinated diphenyl ether , PCDE interference
Journal title
Chemosphere
Serial Year
2002
Journal title
Chemosphere
Record number
736302
Link To Document