Title of article
Micron and sub-micron aerosol sizing with a standard phase-Doppler anemometer
Author/Authors
Gerhard G?bel، نويسنده , , Thomas Wriedt، نويسنده , , Klaus Bauckhage، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 1998
Pages
11
From page
1063
To page
1073
Abstract
In this paper the extended application range of the standard phase-Doppler anemometer (PDA) to particle diameters below the commonly accepted lower limit of about 5–10 μm is described. The problems that accompany measurements on smaller particles are small scattering cross-sections and ambiguous relations between the measured phase difference ΔΦ and the particle diameter d.
A simple, but flexible optical set-up for the transmitting part of a standard PDA system, and a numerical algorithm, which allows careful data evaluation from ambiguous ΔΦ(d) relations, help to manage these problems. Measurements on aqueous suspensions of micron and sub-micron sized dielectric and metalic monospheres, as well as on polydisperse aerosols, are presented and evaluated by means of the proposed algorithm.