Title of article :
Development of a radon-aerosol chamber at NIRS—general design and aerosol performance
Author/Authors :
Hirokazu Ichitsubo، نويسنده , , Yuji Yamada، نويسنده , , Michikuni Shimo، نويسنده , , Akira Koizumi، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
16
From page :
217
To page :
232
Abstract :
A radon-aerosol chamber having an inner volume of approximately 25 m3 has been developed at the National Institute of Radiological Sciences (NIRS). The chamber is designed to control temperature, relative humidity, aerosol concentration, radon concentration, and pressure differentials between inside and outside the chamber. We evaluated its performance as an aerosol chamber. Temperature and relative humidity can be controlled from 278 to 303 K, and 30% to 90%, respectively, and pressure differentials between inside and outside the chamber can be maintained at negative values. Temperature and relative humidity fluctuations are very small. The concentration of particles inside the chamber can be controlled from 108 to 1010 m−3. Particle concentrations and distributions are very stable. The particle concentration fluctuation is within the statistical error of the TSI CPC3010 particle counter. Changes in temperature and relative humidity had little effect on particle concentrations and their distributions.
Keywords :
Radon control , Environmental control , 222Rn , Radon-aerosol chamber , Aerosol particle concentration control