• Title of article

    Sizing of powders in inhalers with an Aerosizer® according to a mixed experimental factorial design

  • Author/Authors

    P. J. Mendes، نويسنده , , A. Raposo، نويسنده , , J. M. M. Sousa، نويسنده , , J. F. Pinto، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    19
  • From page
    509
  • To page
    527
  • Abstract
    In the present work, the Aerosizer® with the Aerodisperser® is used to assess the size and size distribution of two common dry powder inhalers: Ventilan RotacapsTM (salbutamol sulfate) and Bricanyl TurbohalerTM (terbutaline sulfate), according to a mixed factorial design. The different parameters of the equipment (voltage applied to the detector, shear force, deagglomeration, feed rate and pin vibration) were the factors considered in the design, set to different levels. Altogether, a full 41×31×23 factorial was considered and the results of the experiments analyzed according to the Montgomery technique. Results have shown that samples with a single population of particles were less affected by the changes in the parameters, whereas for mixtures incorporating different populations results were affected upon change of the levels of the factors. The voltage at the detector was the most important factor, followed by the shear force. By contrast, pin vibration and feed rate were less significant. The interactions between variables were also responsible for significant changes in the results and should not be neglected throughout the measurements.
  • Keywords
    Aerosizer , Factorial design , Dry-powder inhaler , Montgomery technique , Terbutaline sulfate , Salbutamol sulfate
  • Journal title
    Journal of Aerosol Science
  • Serial Year
    2004
  • Journal title
    Journal of Aerosol Science
  • Record number

    742909