Title of article :
Operation and performance of a differential mobility particle sizer and a TSI 3010 condensation particle counter at stratospheric temperatures and pressures
Author/Authors :
M. Seifert، نويسنده , , R. Tiede، نويسنده , , M. Schnaiter، نويسنده , , C. Linke، نويسنده , , O. Mohler، نويسنده , , U. Schurath، نويسنده , , Linda J. Strom، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Abstract :
A Differential Mobility Particle Sizer (DMPS) system has been adapted and tested for reliable aerosol characterization at stratospheric pressures and temperatures in the AIDA aerosol chamber of Forschungszentrum Karlsruhe. To avoid partial or complete evaporation of volatile aerosol particles (e.g. solution droplets like sulfuric acid droplets), the DMA is operated at almost the same temperature and pressure as the AIDA aerosol vessel. In this study, we describe the experimental setup and technical modifications to the TSI standard DMA 3071 and CPC 3010 used for the DMPS system, discuss the results of low pressure Latex calibration and CPC counting efficiency measurements, and demonstrate the performance of the entire system in a case study with sulfuric acid aerosol.
Keywords :
Latex spheres calibration , DMPS , CPC , Low pressure andtemperature , counting efficiency
Journal title :
Journal of Aerosol Science
Journal title :