• Title of article

    Performance of a venturi dilution chamber for sampling 3–20 nm particles

  • Author/Authors

    Young Jun Yoon، نويسنده , , Sinead Cheevers، نويسنده , , S. Gerard Jennings، نويسنده , , Colin D. O’Dowd، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    535
  • To page
    540
  • Abstract
    The transmission efficiency of a venturi mixing and dilution system was investigated with laboratory generated aerosol by comparison of two condensation particle counters (CPCs). The transmission efficiency exceeded 95% for particle sizes between 3 and 20 nm. The use of the diluter system is demonstrated through comparison with total concentrations derived from a nano-scanning mobility particle sizer (nSMPS) applied to measuring a rapidly changing atmospheric nucleation mode. The study indicates that the diluted-CPC sampler can resolve rapidly changing, and more intense peaks in excess of , which are otherwise missing or under-sampled by the nSMPS.
  • Keywords
    Aerosol dilution
  • Journal title
    Journal of Aerosol Science
  • Serial Year
    2005
  • Journal title
    Journal of Aerosol Science
  • Record number