Title of article
The Aethalometer calibration and determination of iron concentration in dust aerosols
Author/Authors
P. Fialho، نويسنده , , M.C. Freitas، نويسنده , , F. Barata، نويسنده , , B. Vieira، نويسنده , , A.D.A. Hansen، نويسنده , , R.E. Honrath، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2006
Pages
10
From page
1497
To page
1506
Abstract
This paper shows the use of the instrumental neutron activation analysis (INAA) technique to calibrate Aethalometer instruments’ response to iron oxides present in Saharan dust aerosol. The five samples selected for this calibration were collected with a seven-wavelength Aethalometer (model AE31) at the summit of Pico mountain (2225 m ASL) in the Azores Islands. These samples correspond to measurements taken between the 31 October and 5 November, 2001. Analysis of these samples by the INAA technique gave a total mass of (Fe). Correlation of the analytical results with the optical measurements allowed the determination of the elemental iron calibration constant, KFe ( ), which can be used in the determination of iron concentrations from multi-wavelength Aethalometer measurements. As an example of this, we used Aethalometer measurements during the 2001 event to calculate hourly average dust iron, CFe , concentrations, which range from 0.00 and .
Keywords
INAAtechnique , HEMATITE , IRON OXIDES , Aethalometer calibration , Absorption coefficient , Seven-wavelength aethalometer , black carbon , PICO-NARE , Sahara dust
Journal title
Journal of Aerosol Science
Serial Year
2006
Journal title
Journal of Aerosol Science
Record number
743158