Title of article
Correction of detector nonlinearity for the balloonborne Michelson Interferometer for Passive Atmospheric Sounding
Author/Authors
Kleinert، Anne نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
-424
From page
425
To page
0
Abstract
The detectors used in the cryogenic limb-emission sounder MIPAS-B2 (Michelson Interferometer for Passive Atmospheric Sounding) show a nonlinear response, which leads to radiometric errors in the calibrated spectra if the nonlinearity is not taken into account. In the case of emission measurements, the dominant error that arises from the nonlinearity is the changing detector responsivity as the incident photon load changes. The effect of the distortion of a single interferogram can be neglected. A method to characterize the variable responsivity and to correct for this effect is proposed. Furthermore, a detailed error estimation is presented.
Keywords
detectors , CCD , Measurement , Metrology , Optical design of instruments , Optical systems design , instrumentation , Fabrication , Ultrafast measurements , Wave-front sensing , Atmospheric , ocean optics , Ultrafast optics , Optical design , charge-coupled device
Journal title
Applied Optics
Serial Year
2006
Journal title
Applied Optics
Record number
75273
Link To Document