Title of article :
A selective ozone scrubber for application in ambient nitrogen dioxide measurements using the commercial luminoxR (LMA-3, Scintrex/Unisearch Inc.)
Author/Authors :
R. W. H. Schmidt، نويسنده , , J. Kames، نويسنده , , H. J. Kanter، نويسنده , , U. Schurath، نويسنده , , F. Slemr، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
4
From page :
947
To page :
950
Abstract :
A new selective O3 scrubber is described, which effectively removes the complex nonlinear O3 interference: with NO2 measurements made using the commercial LuminoxR detector (LMA-3, Scintrex/Unisearch Inc.). The scrubber, which consists of a small annular denuder coated with natural rubber (1,4-polyisoprene), was optimized for airborne NO2 measurements of 3 to 4 h duration. Ozone is reduced below 1–2 ppb for doses for at least 600 ppb h (ppb: parts per billion (10−9) per volume), while NO2 losses are negligible after preconditioning the scrubber with about 20 ppb h NO2. Simultaneous airborne NO2 measurements in 1992 with two identical LuminoxR instruments showed that the scrubber efficiently removes the systematic overestimation of low tropospheric NO2 mixing ratios by unprotected LuminoxR detectors.
Keywords :
Ozone scrubber , nitrogen dioxide. chemiluminescence , Luminol , Air pollution , measurementtechnique.