• Title of article

    A continuous monitor-sulfur chemiluminescence detector (CM-SCD) system for the measurement of total gaseous sulfur species in air

  • Author/Authors

    Douglas L. MacTaggart، نويسنده , , Sherry O. Farwell، نويسنده , , Julia R. Burdge، نويسنده , , Zhong-Tao Cai، نويسنده , , Timothy J. Haakenson، نويسنده , , W. Lee Bamesberger، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    8
  • From page
    625
  • To page
    632
  • Abstract
    A continuous monitor-sulfur chemiluminescence detector (CM-SCD) system with a flameless, temperature-controlled furnace combustion source was developed for real-time measurement of total sulfur gases in air. This measurement system demonstrated a linear dynamic range exceeding five orders of magnitude and equimolar sensitivity to the most prevalent atmospheric sulfur gases. A detection limit of 10 pptv was obtained using 10 min signal averages. On a real-time basis, detection in the 20–50 pptv range was demonstrated. After modification of its sample inlet system, the CM-SCD showed no appreciable interference effects from the addition of H2O vapor, NO2 or O3 to the sample matrix. In the recent Gas-Phase Sulfur Intercomparison Experiment (GASIE), the CM-SCD compared favorably with SO2 measurement methods.
  • Keywords
    Sulfur chemiluminescence detection , Continuous monitors , Sulfur gases
  • Journal title
    Atmospheric Environment
  • Serial Year
    1998
  • Journal title
    Atmospheric Environment
  • Record number

    755415