Title of article
Direct spectral measurements with a Brewer spectroradiometer: absolute calibration and aerosol optical depth retrieval
Author/Authors
Bais، Alkiviadis F. نويسنده , , Garane، Katerina نويسنده , , Kouremeti، Natalia نويسنده , , Blumthaler، Mario نويسنده , , Kazadzis، Stelios نويسنده , , Gerasopoulos، Evangelos نويسنده , , Schallhart، Barbara نويسنده , , Cede، Alexander نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
-1680
From page
1681
To page
0
Abstract
We present three different methods for the absolute calibration of direct spectral irradiances measured with a Brewer spectroradiometer, which are shown to agree to within +-2%. Direct irradiance spectra derived by Brewer and Bentham spectroradiometers agree to within 4 +- 3%. Good agreement was also found by a comparison of the aerosol optical depth and Angstrom exponent retrieved by the two instruments and a multifilter rotational shadowband radiometer. The spectral aerosol optical depth (300-365 nm) derived from six years of direct irradiance measurements at Thessaloniki shows a distinct seasonal variation, averaging to ~0.3 at 340 nm in winter and ~0.7 in summer.
Keywords
Metrology , Spectrometers , ocean optics , spectroscopic instrumentation , Aerosols , Atmospheric , Measurement , instrumentation , Atmospheric optics
Journal title
Applied Optics
Serial Year
2005
Journal title
Applied Optics
Record number
75646
Link To Document