• Title of article

    Precision of Measurement of Tissue Optical Properties with Optical Coherence Tomography

  • Author/Authors

    Kholodnykh، Alexander I. نويسنده , , Petrova، Irina Y. نويسنده , , Larin، Kirill V. نويسنده , , Motamedi، Massoud نويسنده , , Esenaliev، Rinat O. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -3026
  • From page
    3027
  • To page
    0
  • Abstract
    Accurate and noninvasive measurement of tissue optical properties can be used for biomedical diagnostics and monitoring of tissue analytes. Noninvasive measurement of tissue optical properties (total attenuation and scattering coefficients, optical thickness, etc.) can be performed with the optical coherence tomography (OCT) technique. However, speckle noise substantially deteriorates the accuracy of the measurements with this technique. We studied suppression of speckle noise for accurate measurement of backscattering signal and scattering coefficient with the OCT technique. Our results demonstrate that the precision of measurement of backscattering signals with the OCT technique can be 0.2% for homogeneously scattering media and 0.7% for skin, if spatial averaging of speckle noise is applied. This averaging allows us to achieve the precision of tissue scattering coefficient measurements of approximately -/+0.8%. This precision can be further improved by a factor of 2 -3, upon optimization of OCT operating parameters.
  • Keywords
    aliasing , Consistency , cyclostationary , Doppler , multipath , Spectral density function , harmonizable functions , Estimation , Bias , covariance
  • Journal title
    Applied Optics
  • Serial Year
    2003
  • Journal title
    Applied Optics
  • Record number

    76142