    We previously developed a retinal birefringence scanning (RBS) device to detect eye fixation. The purpose of this study was to determine whether a new binocular RBS (BRBS) instrument can detect simultaneous fixation of both eyes. Control (nonmyopic and myopic) and strabismic subjects were studied by use of BRBS at a fixation distance of 45 cm. Binocularity (the percentage of measurements with bilateral fixation) was determined from the BRBS output. All nonstrabismic subjects with good quality signals had binocularity >75%. Binocularity averaged 5% in four subjects with strabismus (range of 0 -20%). BRBS may potentially be used to screen individuals for abnormal eye alignment.