Title of article :
Combined Twyman-Green and MachZehnder interferometer for microlens testing
Author/Authors :
Zappe، Hans نويسنده , , Reichelt، Stephan نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
A new interferometer design for microlens testing is presented. The phase-shifting system combines the advantages of a Twyman-Green and a Mach-Zehnder interferometer and permits full characterization of the aberrations of microlenses as well as radius of curvature and focal length measurements. The Twyman-Green system is applied to surface testing in reflection (single reflection), whereas the Mach-Zehnder system is used for lens testing in transmission (single pass). Both measurements are performed without removal of the test part, allowing for combination of the results without confusion of the actual lens and without an azimuthal orientation error. The interferometer setup is explained, the test procedure is described, and experimental results are given.
Keywords :
instrumentation , Measurement , Optical instruments , Metrology , Optical design , Metrological instrumentation , Fabrication , Optical testing , Other areas of optics , Micro-optics , Interferometry
Journal title :
Applied Optics
Journal title :
Applied Optics