Title of article
Monitoring of dissolved reactive phosphorus in wastewaters by flow injection analysis. Part 2. On-line monitoring system
Author/Authors
Richard L. Benson، نويسنده , , Yen B. Truong، نويسنده , , Ian D. McKelvie، نويسنده , , Barry T. Hart، نويسنده , , Glenn W. Bryant، نويسنده , , William P. Hilkmann، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
7
From page
1965
To page
1971
Abstract
An automated-flow injection (FI) analyser for the determination of dissolved reactive phosphorus (DRP) in wastewaters is described. The analytical method is based on the formation of molybdenum blue. The limits of detection using a solid-state photometric detector (LED-photodiode) were 0.05 mg P l−1 and the linear range was 0–20 mg P l−1. Details of construction, field-testing (including sample filtration) and validation are given. The analyser proved very reliable during an unattended operational period of seven days. Operational costs were minimised, with low reagent consumption and minimal operator intervention. Data retrieval and analyser status were remotely interrogated by means of a cellular telephone-based communications link. The real-time results from the analyser showed good correlation with those generated by a validated laboratory method.
Keywords
flow injection. phosphorus. wastewaters. on-line monitoring. Filtration
Journal title
Water Research
Serial Year
1996
Journal title
Water Research
Record number
765782
Link To Document