• Title of article

    RESPONSE CHARACTERISTICS OF A DEAD-CELL BOD SENSOR

  • Author/Authors

    Zhenrong Qian، نويسنده , , T.C Tan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    7
  • From page
    801
  • To page
    807
  • Abstract
    Bacillus subtilis cells, thermally killed by exposing dried cells to 280°C for 2.5 mins, were used for fabricating the biofilm attached to a dissolved oxygen probe. The dead cell biosensor showed good sensing characteristics for the BOD of aqueous organic solutions and wastewaters in regard to its response and recovery times, reproducibility and stability. It has a significantly long life span of several months, requiring only simple storage at room temperature in phosphate buffer in between measurements and even over extended idling periods for which microbial sensors would normally require food and nutrient additions for their sustenance. The sensitivity of the biosensor towards 24 organic solutes were generally comparable with biosensors using living Bacillus subtilis cells and the American Public Health Association (APHA) method of measurement using activated sludge. The high diffusion resistance of the membranes was experimentally verified to have caused the long response time for soluble starch. The BOD8 of some real wastewater and synthetic water samples were determined and compared with the BOD5 obtained by the APHA method. The deviation of the biosensor values from the APHA values depends on the difference in the sensitivity of the device and the microbial system towards the compounds in the test sample. Nevertheless the deviation was similar to those generally observed among the APHA values reported for a given substrate by different analysts and laboratories.
  • Keywords
    dead-cell , BOD , Biosensor , Bacillus sub/ilis
  • Journal title
    Water Research
  • Serial Year
    1998
  • Journal title
    Water Research
  • Record number

    766406