Title of article
Colorimetric flow-injection analysis of dissolved iron in high DOC waters
Author/Authors
Michael J. Pullin، نويسنده , , Stephen E. Cabaniss، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
10
From page
363
To page
372
Abstract
An iron flow-injection analysis system has been optimized for the analysis of iron in waters high in dissolved organic carbon. The method detects either dissolved iron(II) or total dissolved iron with a detection limit of 10 nM, precision of 0.65% at 1 μM, and a dynamic range of four orders of magnitude. There are minimal interferences (<1%) from other metals at environmental concentrations. The iron(II) method measures iron(II) in the presence of excess iron(III) with less than 1% interference. When used with pre-acidified samples, the total dissolved iron method agrees well with electrothermal atomic absorption spectrometry for a variety of natural waters with a range of dissolved organic carbon (3–36 mg C/L) and iron (1–28 μM) concentrations. When used with samples at their ambient pH, the total dissolved iron method detects dissolved iron, but not colloidal iron (size fraction 0.05–0.45 μm).
Keywords
colloid , redox , Fresh Water , Iron , Flow-injection analysis , Ferrozine