Title of article :
Simultaneous determination of dissolved organic carbon and total dissolved nitrogen in seawater by high temperature catalytic oxidation: conditions for precise shipboard measurements
Author/Authors :
Xosé A. ?lvarez-Salgado، نويسنده , , Axel E. J. Miller، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
Appropriate conditions have been achieved for the accurate, rapid, and highly precise shipboard simultaneous determination of dissolved organic carbon and total dissolved nitrogen in seawater by high temperature catalytic oxidation. A nitrogen-specific Antek 705D chemiluminescence detector and a CO2-specific LiCor Li6252 IRGA have been coupled in-series with a Shimadzu TOC-5000 organic carbon analyser. Precision of both simultaneous measurements is ≤1.5%, i.e. ±1 μmol C l−1 and ±0.3 μmol N l−1, respectively. Quality of analysis is not compromised by vibrations associated with ocean going research vessels.
Keywords :
DON , Simultaneous measurement , HTCO , DOC
Journal title :
Marine Chemistry
Journal title :