Author/Authors :
Kei Okamura، نويسنده , , Hideshi Kimoto، نويسنده , , Kazuyoshi Saeki، نويسنده , , Junichiro Ishibashi، نويسنده , , Hajime Obata، نويسنده , , Masahiro Maruo، نويسنده , , Toshitaka Gamo، نويسنده , , Eiichiro Nakayama، نويسنده , , YOSHIYUKI NOZAKI، نويسنده ,
Abstract :
This paper presents the first in situ flow-through chemical analyzer using a chemiluminescence (CL) method in the deep sea to a depth of 5200 m. The analyzer, called GAMOS (Geochemical Anomalies MOnitoring System), determines the concentration of dissolved manganese continuously using a H2O2–luminol CL method. A detection limit of 0.23 nM was obtained. The GAMOS was used successfully for hydrothermal plume observation.
Keywords :
manganese , Hydrothermal fields , plumes , chemiluminescence , Flow through , In situ analyzer