Title of article :
Utility of colorimetric end-tidal carbon dioxide detector for monitoring during prehospital cardiopulmonary resuscitation
Author/Authors :
Keiji Nakatani، نويسنده , , Hidekazu Yukioka، نويسنده , , Mitsugu Fujimori، نويسنده , , Chouhei Maeda، نويسنده , , Hiroshi Noguchi، نويسنده , , Shin Ishihara، نويسنده , , Ikuo Yamanaka، نويسنده , , Choichiro Tase، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
4
From page :
203
To page :
206
Abstract :
The purpose of this study was to evaluate a colorimetric end-tidal CO2 ( 2) detector (EASY CAP) as a monitor during prehospital cardiopulmonary resuscitation (CPR) without tracheal intubation. This detector was used for 121 patients during CPR with a laryngeal mask airway or face mask by authorized emergency lifesaving technicians. At 7 to 15 minutes after the initiation of CPR, was <0.5% in 30 cases (group A), 0.5% to 2.0% in 46 cases (group B) and>2.0% in 45 cases (group C). The rate of return of spontaneous circulation was 17% in group A, 24% in group B, and 48% in group C (groups A v C, P< .01). There was a significant difference in the rate of hospital admission between groups A and C. The 2 value may be useful for monitoring during prehospital CPR with a laryngeal mask airway or face mask.
Keywords :
Cardiopulmonary Resuscitation , End-tidal carbon dioxide , out-of-hospital cardiopulmonary arrest , laryngeal mask airway
Journal title :
American Journal of Emergency Medicine
Serial Year :
1999
Journal title :
American Journal of Emergency Medicine
Record number :
779621
Link To Document :
بازگشت