Title of article :
Intraperitoneal blood missed on a FAST examination using portable ultrasound
Author/Authors :
Michael Blaivas، نويسنده , , Daniel Theodoro، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
3
From page :
105
To page :
107
Abstract :
Bedside ultrasonography is rapidly gaining popularity in the evaluation of emergency patients. Several manufacturers have developed hand-held ultrasound machines that make the technology easy to transport and available in settings where such diagnostic tests were previously unobtainable. The miniaturization of equipment often means compromises and no studies exist comparing the high quality imaging capabilities of larger conventional ultrasound units to hand-held machines on actual patients. We present 3 cases in which intra-abdominal fluid stripes, important markers of intraperitoneal bleeding, were not visible with a popular hand-held unit, but were identified with a larger mobile ultrasound machine. These findings should caution emergency physicians to be aware of this limitation along with the many advantages of these new and popular hand-held ultrasound units.
Journal title :
American Journal of Emergency Medicine
Serial Year :
2002
Journal title :
American Journal of Emergency Medicine
Record number :
780189
Link To Document :
بازگشت