Title of article :
The power infuser: A new device for rapid fluid infusion
Author/Authors :
Bernard L. Lopez، نويسنده , , Linda Davis-Moon، نويسنده , , William Sterious، نويسنده , , Michael Bulette، نويسنده , , Xin-liang Ma، نويسنده , , Theodore A. Christopher، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
4
From page :
129
To page :
132
Abstract :
This study examined the effectiveness and ease of use of the ID Power Infuser, a pocket-sized device that can infuse fluid at rates up to 6 L/hr. Forty-six adults presenting with non-traumatic hypotension or clinical hypovolemia had 0.9% NaCl solution infused by the ID Power Infuser through a peripheral IV catheter. Eighteen patients with hemodynamic evidence of hypovolemia had a mean infusion rate of 2 L/hr with time to resolution of hypovolemia 28.5 ± 7 min, 14 of these had an infusion rate ≥ 4 L/hr with resolution of hypovolemia of 13.7 ± 5.8 min. Resolution of hypovolemia was significantly faster compared with a group receiving gravity infusion. Use of the Power Infuser was deemed easy (VAS = 1.9 ± 1.9 cm). The ID PowerInfuser can accurately and quickly deliver IV crystalloid in the ED and has the potential to reduce morbidity, time in the ED, and costs.
Journal title :
American Journal of Emergency Medicine
Serial Year :
2003
Journal title :
American Journal of Emergency Medicine
Record number :
780300
Link To Document :
بازگشت