Title of article
Design of coils for millimeter- and submillimeter-sized biotelemetry
Author/Authors
S.F.، Pichorim, نويسنده , , P.J.، Abatti, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
-1486
From page
1487
To page
0
Abstract
Biotelemetric systems, especially those that employ implanted devices, work with inductive links, where usually large circular external coils are separated by relatively large distances (dimensions of centimeters) from the small (millimeter- or submillimetersize) implanted coils. This paper shows that, under these conditions, a simplified method for calculation of the mutual inductance (M) between the coils, avoiding elliptic integrals, can be obtained. A procedure for coil design, with maximum M between them, is also described.
Journal title
IEEE Transactions on Biomedical Engineering
Serial Year
2004
Journal title
IEEE Transactions on Biomedical Engineering
Record number
80526
Link To Document