Title of article :
A new colour M-mode index of diastolic filling compared with radionuclide ventriculography
Author/Authors :
Torkel Steen، نويسنده , , Kjell Rootwelt، نويسنده , , Cecilie Ris?e، نويسنده , , Arne Kristian Andreassen، نويسنده , , Halfdan Ihlen، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
7
From page :
89
To page :
95
Abstract :
We correlated the new diastolic index ‘delay of apical peak velocity’, as measured by colour M-mode Doppler, with radionuclide ventriculographic indices of ventricular function. Thirty-seven patients with coronary artery disease participated in the prospective and blinded study, which included repeated acquisitions to determine the effect of realigning the Doppler sample beam. In multiple regression, neither peak filling rate, left ventricular phase histogram width nor ejection fraction were statistically significantly related to delay of apical peak velocity. The standard deviation of the differences between duplicate colour M-mode acquisitions corresponded to half the reference range of the index. We conclude that in this blinded investigation, the new Doppler index did not provide information about ventricular function equivalent to radionuclide ventriculography. The index may be significantly influenced by sample beam position.
Keywords :
Colour M-mode echocardiography , Diastole , Doppler ultrasound , radionuclide ventriculography , Ventricularfunction
Journal title :
International Journal of Cardiology
Serial Year :
1995
Journal title :
International Journal of Cardiology
Record number :
811809
Link To Document :
بازگشت