Title of article :
Lost and found: An unusual late complication of the Angio-Seal closure device
Author/Authors :
Felicita Andreotti، نويسنده , , Alberto Lavorgna، نويسنده , , Giulio Coluzzi، نويسنده , , Claudio Vincenzoni، نويسنده , , Antonio G. Rebuzzi، نويسنده , , Italo Porto، نويسنده , , Filippo Crea، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
We report an unusual case of breakage and embolisation of a femoral access closure device (Angio-Seal), which caused acute leg ischemia after percutaneous coronary intervention in a diabetic patient with documented peripheral vascular disease. Embolectomy from the superficial femoral artery revealed the polymer anchor split in two parts and embedded within a large (25 × 5 mm) thrombus.
Keywords :
Breakage of vascular closure device , Acute leg ischemia , peripheral artery disease
Journal title :
International Journal of Cardiology
Journal title :
International Journal of Cardiology